Glossary
Terms of art across modalities, in plain language.
156 entries
5-gauss line
The boundary around an MRI magnet where the fringe field drops below 5 Gauss (0.5 mT). Below 5 Gauss, the field is generally considered…
510(k)
FDA's premarket pathway for devices substantially equivalent to existing legally marketed predicates. Cleared devices receive a "K-number"…
a-Se detector
Amorphous selenium flat-panel detector. Direct-conversion X-ray detector — selenium converts incident X-ray photons directly to electrical…
ACR accreditation
American College of Radiology accreditation for imaging facilities. Required (or near-required by payer policy) for CMS reimbursement of…
ADMIRE
Siemens Advanced Modeled Iterative Reconstruction — the model-based iterative successor to SAFIRE on Siemens CT platforms. Standard on…
AE title
Application Entity title — DICOM identifier for a network-connected device. Used in every DICOM association handshake (C-ECHO, C-STORE,…
Affirm biopsy
Hologic's stereotactic biopsy attachment. Converts a Selenia Dimensions / 3Dimensions mammography unit into an upright stereotactic-biopsy…
Agility MLC
Elekta's current-generation 160-leaf multileaf collimator. Tungsten leaves with a projected width of 5 mm at isocenter across the full…
AiCE
Canon Medical's Advanced intelligent Clear-IQ Engine — deep-learning CT reconstruction. Successor to the iterative-reconstruction-era AIDR…
AIDR 3D
Canon Medical (legacy Toshiba) Adaptive Iterative Dose Reduction 3D — the first-generation iterative reconstruction available across the…
AIR Coils
GE HealthCare's blanket-like conforming receive coils for MRI. The AIR coil platform replaces rigid body / spine / head coils with…
AIR Recon DL
GE HealthCare's deep-learning MRI reconstruction — the defining software feature on the current Signa MRI family (Artist 1.5T, Explorer,…
anode heat capacity
Total thermal energy an X-ray tube's anode can absorb before reaching damage threshold. Measured in HU (heat units) or MHU (mega-heat…
ASiR
GE's first-generation iterative reconstruction algorithm. Reduces image noise, enabling lower-dose scans at equivalent image quality.…
ASiR-V
GE's second-generation model-based iterative reconstruction. Successor to ASiR; standard on Revolution CT-era platforms and…
B-mode
Standard 2D grayscale ultrasound image. 'B' for brightness — pixel intensity represents the amplitude of the returning echo at each depth…
BGO
Bismuth germanate (Bi₄Ge₃O₁₂) scintillator. Used in pre-ToF PET (Discovery VCT, Discovery STE). High light yield but slow decay (~300 ns)…
BioMatrix
Siemens Healthineers' BioMatrix Technology — a platform-level MRI feature set introduced with the MAGNETOM Vida (first 3T BioMatrix, ECR…
C-View
Hologic's synthesized 2D image, reconstructed from the 3D tomosynthesis projection set rather than a separate 2D exposure. FDA-approved…
CBCT (dental)
Dental cone-beam CT — volumetric imaging for dental implants, endodontics (root-canal anatomy), orthognathic / orthodontic planning, TMJ…
CBCT (rad onc)
Cone-Beam CT — volumetric kV imaging acquired on the linac's on-board imaging system before each treatment fraction. Standard for…
cephalometric
Lateral skull radiograph used in orthodontic planning. Standardized geometry — patient positioned in a cephalostat with fixed…
CertaRay
Philips's late-generation Allura Xper generator (paired with R8.2 software). Upgradeable to Azurion HE — the software / workstation /…
ClarityIQ
Philips's image-processing and acquisition pipeline for interventional X-ray that reduces fluoroscopic dose 43–83% versus baseline Xper…
cold head
Two-stage cryocooler that maintains the MRI magnet's cold shields at ~40 K, limiting helium boil-off from the inner cryogen reservoir at…
collimation
Restricting the X-ray beam to the region of interest. In CT, collimation determines beam width along the z-axis (e.g., 64 × 0.625 mm = ~40…
Color Doppler
Doppler imaging mode that displays blood-flow velocity as a colored overlay on the B-mode image. Convention: red toward the probe, blue…
Compressed SENSE
Philips' combination of compressed sensing + SENSE parallel imaging — MRI acceleration that can shorten acquisition times by up to ~50% on…
compression paddle
Clear plastic mammography compression paddle that flattens breast tissue against the detector for imaging. Compression evens tissue…
coneXact
Canon Medical (legacy Toshiba) coneXact reconstruction algorithm — the software that enables the Aquilion CX / CXL to generate 128 unique…
CR
Computed Radiography — legacy phosphor-plate-and-reader system that bridged film-screen and direct DR. Photostimulable phosphor plate…
CTDI
Computed Tomography Dose Index — standardized measurement of CT dose per scan rotation, measured in a CTDI phantom. Reported in mGy. The…
curved array
Curved-footprint (convex) ultrasound probe for abdominal, OB / GYN, and general radiology imaging. Wider field of view than linear, at…
CW Doppler
Continuous-wave Doppler — measures peak velocities beyond PW Doppler aliasing limits. The trade-off: no range-gate, so the velocity…
DAP
Dose-Area Product (also kerma-area product, KAP) — fluoroscopic dose multiplied by beam area at the patient surface, integrated over the…
DAS
Data Acquisition System. The electronics chain that digitizes the analog signal from the CT detector array — converting per-channel…
De Novo
FDA pathway for novel low / moderate-risk devices without a predicate. Briefer than PMA but more rigorous than 510(k) — establishes a new…
Deep Resolve
Siemens Healthineers' deep-learning MRI reconstruction — accelerates acquisition and/or raises effective image quality on…
DICOM
Digital Imaging and Communications in Medicine — the universal standard for medical-imaging file formats and network communication. Every…
DICOM MWL
DICOM Modality Worklist — pulls scheduled exams from RIS / EHR to the scanner console, eliminating manual demographic entry at the…
DICOM Store
DICOM C-STORE transaction — scanner pushes images to PACS over a DICOM association. The basic "send images to archive" operation, the…
DLIR / TrueFidelity
GE's deep-learning CT image reconstruction — branded TrueFidelity, commonly referred to by the generic DLIR (deep-learning image…
DLP
Dose-Length Product — total CT dose × scan length, expressed in mGy·cm. Summary dose metric for a CT exam, derived from CTDI and reported…
DQE
Detective Quantum Efficiency — measure of how efficiently an X-ray detector converts incoming photons into useful signal. Expressed as a…
DR panel
Direct (digital) radiography flat-panel detector — wired or wireless, with CsI on a-Si (indirect-conversion scintillator) or a-Se…
dS Coils
Philips digital MR receive coils for the dStream broadband architecture. RF digitization is performed at the coil connector; optical /…
dStream
Philips' digital receive architecture for MRI. Analog-to-digital conversion moves from the scanner console out to the coil itself — the…
DWI
Diffusion-Weighted Imaging — MR sequence sensitive to the random Brownian motion of water molecules. Areas of restricted diffusion (acute…
endocavity
Small-diameter curved or microconvex ultrasound probe designed for internal-cavity imaging — transvaginal (TV), transrectal (TR), and…
EPI
Echo-Planar Imaging — ultra-fast MR sequence that acquires the whole 2D image in one excitation by rapid gradient-reversal switching.…
F-18 FDG
Fluorine-18 fluorodeoxyglucose — the dominant PET oncology tracer. Glucose analog labeled with the F-18 positron emitter. Half-life ~110…
FDA recall class
FDA classifies device recalls into three classes by severity:
FDG (fluorodeoxyglucose)
Fluorine-18-labeled fluorodeoxyglucose — the dominant PET tracer globally. Glucose analogue that is taken up by metabolically active cells…
FFDM
Full-Field Digital Mammography — digital 2D mammography. Replaced screen-film mammography as the reimbursed standard in the 2000s.…
FFF (Flattening-Filter-Free)
Linac beam mode where the flattening filter is removed from the treatment head. Profile is forward-peaked instead of flat, but dose rate…
FHIR
Fast Healthcare Interoperability Resources — modern RESTful healthcare interoperability standard. JSON-based, web-native. The HL7…
filtered back projection
Classical CT reconstruction algorithm — analytic single-pass method that filters projection data and back-projects to form the image. The…
FLAIR
Fluid-Attenuated Inversion Recovery — MR inversion-recovery sequence that nulls cerebrospinal-fluid signal. Periventricular lesions…
fluoroscopy
Real-time X-ray imaging during procedures. Continuous-wave or pulsed acquisition with the operator visualizing anatomy and instruments…
focal spot
The effective area of the anode where X-rays are produced. Smaller focal spots produce sharper images but dissipate heat less efficiently…
gantry aperture
Diameter of the opening in the CT gantry through which the patient table passes. 70 cm is the standard clinical bore; 78–82 cm wide-bore…
Ge-68
Germanium-68 rod source used for PET daily QC and normalization calibration. Half-life ~271 days. Decays to Ga-68 (positron-emitting…
Gemstone detector
GE HealthCare's garnet-based CT scintillator — the first new commercial CT scintillator material in roughly 20 years at launch. The…
gradient strength
Magnetic field gradient amplitude, measured in mT/m. Defines how sharply the field varies across space — combined with slew rate (T/m/s),…
GSI (Gemstone Spectral Imaging)
GE HealthCare's fast-kVp-switching dual-energy CT approach. The X-ray tube rapidly alternates between low-energy (typically 40 kVp) and…
helical
CT acquisition mode where the patient table moves continuously through the gantry while the gantry rotates and the X-ray tube fires.…
HL7
Health Level Seven — legacy healthcare messaging standard (v2.x) for clinical and administrative data interchange. The dominant interface…
HU
Hounsfield Unit — the CT image pixel value, normalized to water (0 HU) and air (−1000 HU). Named for Godfrey Hounsfield, who built the…
HVAC load
Heat-rejection requirement placed on the room's air-conditioning by imaging equipment. Measured in kW or BTU/hr. Comprises cabinet heat…
IEC 60601
International electrical-safety standard for medical electrical equipment. Modality-specific particular standards include 60601-2-33…
IHE
Integrating the Healthcare Enterprise — non-standards-body that defines implementation profiles for healthcare interoperability. IHE…
IMRT
Intensity-Modulated Radiation Therapy — uses MLC to vary beam intensity across each treatment field, producing highly conformal dose…
iPAT
Siemens's integrated Parallel Acquisition Technique — parallel-imaging acceleration that uses multi-coil spatial information to skip…
iterative reconstruction
Family of CT reconstruction methods that iteratively refine the image to reduce noise at the cost of more computation per slice. Modeled…
klystron
High-power microwave vacuum-tube amplifier. Generates the RF energy that accelerates electrons through the linac waveguide. Used on…
kVp
Peak kilovoltage across the X-ray tube. Sets the energy spectrum of the X-ray photons; together with mA·s (tube current × exposure time),…
linear array
Flat-footprint ultrasound probe with parallel elements. High-frequency (5–18 MHz typical) — short-range, fine-detail imaging at the cost…
LYSO
Cerium-doped lutetium-yttrium orthosilicate scintillator (Lu(2-x)YxSiO5:Ce). Fast decay time (~40 ns) and high light yield make it the…
M-mode
Ultrasound acquisition along a single sample line, plotted over time. The result is a 1D position × time strip — high temporal resolution…
mA
Tube current in milliamperes. Controls the number of X-ray photons produced per unit time. Together with kVp (peak voltage) and exposure…
mA·s
Product of tube current (mA) and exposure time (seconds). The primary exposure parameter in radiography and the core dose driver. CT…
magnetron
Simpler RF source than the klystron — a self-oscillating microwave generator that produces the RF energy for the linac waveguide. Used on…
matrix array
Ultrasound transducer with a 2D grid of elements — enables electronic volumetric (3D / 4D) beam steering without mechanical scanheads.…
MAUDE
Manufacturer and User Facility Device Experience — the FDA adverse-event reporting database. Manufacturers must submit MDRs (Medical…
MHU
Mega Heat Units — the thermal capacity of an X-ray tube's anode. 1 HU = 1 watt-second × 1.35 (for three-phase generators); 1 MHU = 1…
Millennium 120 MLC
Varian's standard-tier multi-leaf collimator. 120 leaves total, 5 mm leaf width at isocenter across the central 20 cm field, 10 mm at the…
MLC
Multi-Leaf Collimator — array of motorized tungsten leaves that dynamically shape the treatment beam aperture, enabling IMRT, VMAT, SRS /…
MPPS
Modality Performed Procedure Step — DICOM message from the modality to RIS announcing procedure start / complete / discontinue. Closes the…
MQSA
Mammography Quality Standards Act — federal law requiring FDA certification, ACR (or approved state body) accreditation, annual…
MRC tube
Philips's family of high-output rotating-anode X-ray tubes used in cath-lab and interventional systems. The dominant tube class on the…
MU
Monitor Unit — the unit of linac dose-delivery measurement, defined by the dose registered on the in-head ionization chamber under a…
MU/min
Monitor Units delivered per minute — linac dose rate. Standard tiers:
NRC
Nuclear Regulatory Commission — federal U.S. regulator of byproduct radioactive material in medicine. Agreement States (37 states)…
nSIGHT Imaging
Philips' parallel-pixel ultrasound beamformer architecture on the EPIQ premium platform. Replaces the legacy line-by-line scan-converter…
OBI
On-Board Imager — Varian's on-linac kV imaging system. A kV X-ray tube and flat-panel detector are mounted on retractable arms orthogonal…
PACS
Picture Archiving and Communication System — the hospital's image archive and viewing infrastructure. Every modality's images flow into…
panoramic X-ray
Single curved 2D dental image showing all teeth and jaw anatomy in one acquisition. The X-ray tube and detector rotate around the patient…
Performix Pro
GE's high-output CT X-ray tube family. Performix Pro VCT 100 (~8.0 MHU) is the LightSpeed VCT-class tube. Predecessor and value-tier…
PET
Positron Emission Tomography — detects coincident 511 keV photons from positron annihilation following injection of a positron-emitting…
phantom
Physical test object with known properties used to verify imaging scanner calibration. Modality-specific:
phased array
Small-footprint ultrasound probe with closely-spaced elements driven with phase-controlled timing to electronically steer the beam.…
Photon-Counting CT
A category of CT detector technology that counts and energy-resolves individual X-ray photons rather than integrating their summed signal…
Pianissimo
Canon Medical (legacy Toshiba) Pianissimo Noise Reduction — the MRI gradient acoustic-noise-reduction technology that shipped on the…
pitch
In helical CT, the ratio of table travel per rotation to total detector z-axis coverage. Pitch 1 spreads the helix thinner with gaps;…
PMA
FDA Premarket Approval — full safety and effectiveness review for high-risk or novel devices. The most rigorous FDA pathway, distinct from…
PMT
Photomultiplier Tube — vacuum-tube photodetector that converts scintillation light from gamma-camera or PET crystals into an electrical…
pre-scan
Low-dose localizer image acquired before the diagnostic CT scan. The X-ray tube parks in a fixed position (AP / lateral) while the table…
probe port
Physical receptacle on an ultrasound cart where a transducer connects. Premium carts have 3–4 active probe ports for rapid swapping…
pulse sequence
A programmed timing pattern of RF pulses and gradient switching that acquires MRI images. T1, T2, FLAIR, DWI, EPI, gradient-echo, MR…
Pulsed Wave
Pulsed-wave Doppler — Doppler sampling from a defined range gate along the beam, measuring velocity at a specific anatomical point. The…
PureWave
Philips's single-crystal ultrasound transducer technology. Replaces the conventional polycrystalline-ceramic piezoelectric stack with a…
Q.Clear
GE HealthCare's Bayesian penalized-likelihood PET reconstruction algorithm. Combines maximum-likelihood iterative reconstruction with a…
QMP
Qualified Medical Physicist — board-certified clinical medical physicist (American Board of Radiology, American Board of Medical Physics,…
QuantaMax
Siemens' photon-counting CT detector — the direct-conversion cadmium-telluride (CdTe) detector at the heart of the NAEOTOM Alpha family.…
quench
Catastrophic loss of superconductivity in the magnet. The superconducting wire transitions to normal resistance, energy stored in the…
RapidArc
Varian's volumetric-modulated arc therapy (VMAT) implementation. During beam delivery, the gantry rotates continuously while the multileaf…
reconstruction kernel
Mathematical filter applied during CT image reconstruction. Sharp kernels (bone, lung) emphasize edges and high-frequency content at the…
RF amplifier
Radiofrequency Power Amplifier (RPA) — broadband, high-power amplifier that drives the MRI transmit coil. Duty-cycle-limited; produces the…
RIS
Radiology Information System — scheduling, ordering, and reporting system for radiology. Typically integrated with PACS (image storage /…
rotation time
Time for one full gantry rotation on a CT scanner. Typical range: 0.25–0.35 s (fast, cardiac-tier) to 2.0+ s (slow, low-dose pediatric /…
RSO
Radiation Safety Officer — the named, qualified individual responsible for the radiation-safety program at a facility that handles…
rubidium-82
Rb-82 — cardiac PET perfusion tracer. Generator-produced from strontium-82 parent (~25-day half-life in generator); very short half-life…
SAR
Specific Absorption Rate — RF power deposited in tissue per unit mass, measured in W/kg. Regulatory limit protects against tissue heating…
SBRT
Stereotactic Body Radiotherapy — SRS applied outside the cranium. Lung, liver, prostate, spine, pancreas, adrenal, oligometastatic…
SENSE (SENSitivity Encoding)
Philips' parallel imaging technique for MRI. Uses the spatial sensitivity profiles of a multi-element receive coil to reconstruct the full…
shim
Process of making the MRI magnetic field homogeneous across the scan volume. Two layers:
SieClear
Siemens' spatial-compounding ultrasound imaging technology. Acquires multiple steered-angle scan lines per frame (up to 13 lines of sight…
SiPM (silicon photomultiplier)
Solid-state semiconductor light-sensor technology that replaces conventional photomultiplier tubes (PMTs) in modern PET detectors. A SiPM…
slew rate
Rate at which the MR gradient field changes, measured in T/m/s. Defines how fast the gradient can switch between values within a pulse…
slice count
Number of simultaneous slices a CT scanner acquires per gantry rotation. Drives whole-body coverage speed, cardiac temporal resolution,…
slip ring
Rotating electrical interface in a CT gantry. Transfers power, control signals, and detector data between the stationary frame and…
SnapShot Pulse
GE's prospective ECG-gated cardiac CT acquisition mode. Synchronizes scan timing to a single cardiac phase per heartbeat, eliminating the…
SPECT
Single-Photon Emission Computed Tomography — gamma camera rotates around the patient acquiring projection data, reconstructed into 3D…
SRS
Stereotactic Radiosurgery — single-fraction (or few-fraction, "fractionated SRS") high-precision ablative radiation. Brain metastases,…
STRATON
Siemens' rotating-anode X-ray tube family with a liquid-metal bearing instead of the conventional ball-bearing anode axle. A thin film of…
T-score
Bone density expressed in standard deviations relative to young-adult peak bone mass. The headline DXA clinical number used in WHO…
Tesla
Unit of magnetic field strength. 1 T = 10,000 Gauss. Named for Nikola Tesla. Clinical MRI field strengths span:
TG-142
AAPM Task Group 142 — the consensus standard for linear-accelerator quality assurance. Defines daily, monthly, and annual test schedules.…
thyratron
High-voltage switching vacuum tube in the linac's pulse-forming network (PFN). Drives the modulated high-voltage pulse that feeds the…
TIM
Siemens's Total imaging Matrix coil platform — multi-element matrix coils that can combine up to dozens of elements across body coverage…
Tim 4G
Siemens' fourth-generation Total imaging matrix coil and receive platform. Introduced on MAGNETOM Aera / Skyra (2010) and carried across…
ToF
Time-of-Flight PET — measures the timing difference between the two coincident 511 keV photons from a single annihilation event,…
tomosynthesis
Digital Breast Tomosynthesis (DBT) — the X-ray tube sweeps a limited arc (typically ±7.5° to ±25° depending on vendor) acquiring ~9–25…
transducer
Ultrasound probe — converts electrical signal to acoustic waves and back. Piezoelectric crystal stacks (conventional) or silicon CMUT…
TrueFidelity DLIR (GE Deep-Learning CT Reconstruction)
GE HealthCare's deep-learning image reconstruction stack for CT — the FDA-cleared (2019) trained-neural-network reconstruction algorithm…
TrueV
Siemens' extended-axial-FOV PET detector option on Biograph TruePoint-era platforms (Biograph 16 / 40 / 64). TrueV adds 48 detector blocks…
tube life
Expected useful life of an X-ray tube, typically measured in scans, mA·s, or hours depending on modality. The single most-watched…
UFC detector
Siemens' Ultra Fast Ceramic scintillator — the CT detector material that shipped on the SOMATOM Sensation family and continued into…
Velara
Philips's earlier high-frequency interventional-X-ray generator platform — revisions R7.x and R8.1 shipped on Allura Xper. Predecessor to…
VMAT
Volumetric Modulated Arc Therapy — the gantry rotates continuously while MLC, dose rate, and gantry speed modulate simultaneously.…
Volume Shuttle
GE's table-shuttle acquisition mode that covers volumes larger than the detector z-coverage in a single breath-hold by oscillating the…
waveguide
Copper cavity structure that accelerates electrons through cavity-by-cavity RF coupling driven by the klystron or magnetron.…
window / level
Display parameters that map CT Hounsfield Unit values to screen brightness. "Level" centers the display range; "window" sets the range…
xMATRIX
Philips matrix-array transducer technology. The flagship X6-1 probe uses ~9,000 active elements — roughly 35× a conventional phased-array…
Z-score
Bone density expressed in standard deviations relative to age-matched peers (versus T-score, which compares to young-adult peak). Z ≤ −2.0…
z-Sharp
Siemens' flying focal spot technology. Inside the STRATON tube, an electron-beam deflection mechanism rapidly switches the focal spot…