Siemens Biograph 64 (TruePoint / TrueV)
Family: Siemens Biograph TruePoint · Modality: PET/CT
Flagship of the Siemens Biograph TruePoint family and direct competitor-era PET/CT to the GE Discovery 690 / Discovery STE / VCT class. Biograph 64 pairs the LSO PET detector architecture with a 64-slice CT derived from the SOMATOM Sensation 64 platform — bringing z-Sharp flying focal spot, STRATON tube, 80 kW generator, and CARE Dose4D into an integrated PET/CT chassis. Large global installed base; one of the most-traded Siemens premium PET/CT platforms in the secondary market.
PET detector architecture (TruePoint / TrueV)
- LSO crystals — 4.0 × 4.0 × 20 mm; 169 crystals per detector block.
- 144 detector blocks on base configuration; 192 blocks with TrueV option.
- Transaxial FOV 605 mm.
- Axial FOV — 162 mm base; 216 mm with TrueV.
- 81 image planes base; 109 with TrueV.
- ToF performance — ~540 ps coincidence timing on ToF-capable revisions.
- NECR peak — 96 kcps at 35 kBq/cc base; 165 kcps at 32 kBq/cc with TrueV.
- Sensitivity — 4.2 cps/kBq at 435 keV base; 7.6 cps/kBq with TrueV.
- 435 keV energy threshold — tight energy window enabled by LSO's high energy resolution.
TrueV option
The TrueV configuration adds 48 detector blocks, extending the axial FOV from 162 mm to 216 mm. Clinical consequences:
- Fewer bed positions per whole-body scan — shorter total acquisition time or higher sensitivity at matched time.
- Higher peak NECR — 165 kcps vs 96 kcps base; supports higher-activity protocols or lower injected dose at matched image quality.
- Standard on most late-production Biograph 64 units; verify at acceptance.
CT side (Sensation 64 lineage)
- 64 reconstructed slices via z-Sharp flying focal spot on a 32-row UFC detector (shared with Sensation 64 standalone CT).
- 0.33 s minimum rotation — cardiac CTA workflows supported.
- STRATON tube · 80 kW high-frequency generator · 5 MHU/min heat dissipation.
- CARE Dose4D — 4D tube-current modulation.
- syngo CT host.
Specs
- LSO PET (4.0 × 4.0 × 20 mm crystals)
- Transaxial FOV 605 mm · Axial FOV 162 mm / 216 mm TrueV
- 144 detector blocks / 192 TrueV
- ToF ~540 ps · NECR 96 / 165 kcps · Sensitivity 4.2 / 7.6 cps/kBq
- 64-slice CT (Sensation 64 lineage) · STRATON tube · 80 kW generator · z-Sharp
- syngo CT + PET host
Parts
- Siemens LSO PET Detector Block
- Siemens STRATON X-ray Tube (CT side)
- Siemens UFC Detector Module (CT side)
Clinical positioning
- Premium oncology PET/CT — FDG staging, restaging, response imaging at top-tier image quality for its era.
- Cardiac PET/CT — diagnostic CT coronary calcium / CTA overlaid with Rb-82 myocardial perfusion PET.
- Whole-body staging with diagnostic CT — 64-slice CT supports diagnostic-quality CT in a single exam, eliminating separate diagnostic CT workflow.
- Neuro PET — amyloid, FDG dementia, neuroendocrine.
- Theranostic imaging — Ga-68 DOTATATE, Ga-68 PSMA workflows.
Market position (secondary / refurb)
- Large global install base. Biograph 64 is one of the most-traded premium PET/CT platforms in the secondary market, alongside Discovery 690 and Gemini TF.
- Mature service ecosystem — Siemens OEM service; multi-vendor service coverage on LSO detector and STRATON CT tube.
- Successor pathway — Biograph mCT (analog ToF with Extended Axial FOV) → Biograph Vision (digital SiPM) → Biograph Vision.X (current flagship).
Relationship to siblings
- Biograph 16 — 16-slice entry-tier sibling.
- Biograph 40 — 40-slice intermediate sibling.
- Biograph Duo — 2-slice predecessor-class.
- Biograph mCT — direct successor (analog ToF).
- Biograph Vision / Vision.X — digital SiPM successors.