Philips Allura Xper Family

Fixed interventional X-ray platform that anchored cath / angio / IR rooms for 13 years. Multiple distinct configurations — single-plane cardiac (FD10), single-plane angio/vascular/neuro (FD20), biplane (FD10/20), and the dose-reduction ClarityIQ-equipped Allura Clarity FD20.

Succeeded by the Azurion platform (ClarityIQ built in; ultra-fast workflow). Allura Xper units remain active in many rooms — Philips offers direct upgrade paths from R8.2 to Azurion HE for centers that want to stay on current software without replacing the hardware.

Variants (each a separate System Card)

Shared specs

  • Ceiling-suspended gantry
  • 100 kW high-frequency X-ray generator (Velara R7.x / R8.1 · CertaRay R8.2)
  • MRC-class high-output X-ray tube
  • CU 3000 cooling unit (direct continuous forced-oil cooling)
  • Five functional blocks: Geometry · X-ray Generation · User Interface · Image Detection · Viewing
  • Hemodynamics integration (Xper IM)
  • Software releases: R7.x → R8.1 → R8.2 (upgradeable to Azurion HE)

Shared install

Fixed cath-lab build-out — 3-phase 480V ~150 kVA, ceiling reinforcement for gantry load, lead-shielded room, hemodynamics cabling, sterile workflow plumbing, HVAC. 2–4 week install plus weeks of room build-out.

Shared field guide

See field-guide/philips-allura-xper — MRC tube life, flat-detector pixel integrity, generator revision compatibility (Velara vs CertaRay), cooling loop discipline, gantry geometry calibration.